Neomedica’s well-established 5-diff cell counter – an open system with the addition of reticulocyte analysis forms an important constituent of any middle-sized laboratory. Phoenix NCC-51 is an important part of Neomedica’s 5-diff portfolio and has already proven itself as a gold standard for comprehensive results in labs and hospitals of moderate workload and throughput. Alongside only 4 onboard reagents (1 sheath), this analyzer uses a separate reagent for reticulocyte analysis as an additional benefit to medical practitioners.
With 4 different determination methods and through calculations of different angles via laser scatters, this instrument delivers the complete picture, including cell size, granularity and complexity. Characterized by simplicity in operation, a comprehensive QC program and high-quality reagents, this robust device serves to improve the quality of healthcare and to reduce the costs by minimizing the consumption and providing fast, reliable and accurate results at the same time.
Technical specification
Principle | WBC/DIFF: Flow cytometry, semiconductor, laser light multi-dimensional cell classification WBC Analysis: Optical and impedance measurementsRBC/PLT Analysis: Impedance MethodHGB Test: Cyanide-free reagent colorimetry |
Working mode | Double WBC counting mode (impedance & optical) |
5-part differentiation of WBC
28 parameters + 2 histograms for RBC and PLT, 2 5-part differential scatter-grams for eosinophils and neutrophils and 2 3-D stereograms |
|
Parameters | WBC, LYM#, MON#, NEU#, EOS#, BAS#, LYM%, MON%, NEU%, EOS%, BAS% , RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR, P-LCC
3 RETIC parameters: RETIC, RETIC_ABS, IRF |
Mode | Whole blood, Prediluted |
Sample volume | Whole blood 20 µl, Prediluted 20 µl |
Dimension of aperture | WBC: 100 µm, RBC/PLT: 68 µm |
General throughput | Up to 60 samples/hour with 50 autoloader positions |
Storage | Up to 200.000 sample results including histograms |
QC mode | LJ, X, XR, XB etc. |
Calibration mode | 3 calibration modes: automatic calibration, whole blood calibration and manual calibration |
Windows operating system | |
Display | PC display (14 inch color TFT) |
Power supply | 220V~50/60 Hz |
Interface | Bidirectional LIS and barcode reading |
Report | External printer |
Automatic reagent residual alarm, different warning alarms announce probable abnormalities of the specimen | |
Dimensions | 660x560x545mm (WxDxH), 73 kg |